SE-601 Series Electrocardiograph User Manual
Guidance and manufacture's declaration - electromagnetic immunity -
Guidance and manufacture's declaration - electromagnetic immunity
The SE-601 Electrocardiograph is intended for use in the electromagnetic environment
specified below. The customer or the user of SE-601 Electrocardiograph should assure that it is
used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Power frequency
(50Hz/60Hz)
magnetic field
IEC/EN 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
for all EQUIPMENT and SYSTEMS
IEC/EN 60601 test
level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV line to line
2 kV line to ground
3A/m
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
Compliance level
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV line to line
2 kV line to
ground
3A/m
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 sec
- 120 -
EMC Information
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic
material,
relative humidity should be
at least 30%.
Mains
power
should be that of a typical
commercial
or
hospital
environment.
Mains
power
should be that of a typical
commercial
or
hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
commercial
or
hospital
environment.
Mains
power
should be that of a typical
commercial
or
hospital
environment. If the user of
the SE-601
Electrocardiograph
requires
continued
operation
during
mains interruptions, it is
recommended
that
SE-601
Electrocardiograph
powered
from
uninterruptible
supply or a battery.
the
quality
quality
typical
quality
power
the
be
an
power