Configuring A Modality Worklist Server - Siemens ACUSON Freestyle User Manual

Diagnostic ultrasound system
Hide thumbs Also See for ACUSON Freestyle:
Table of Contents

Advertisement

Configuring a Modality Worklist Server

To configure a DICOM Modality Worklist Server:
1.
Go to
/
SETUP
SETTINGS
2.
Select
.
NEW
3.
Enter the DICOM Worklist information as described in the following table.
DICOM Worklist Fields
Field
Description
Alias
Name you define for the server.
Limited to 16 characters.
IP Address
The IP address of the Server.
AE Title
Identifies the server application that receives worklist messages from
the
field must exactly match (including case) the Application Entity name
assigned to the server application.
Limited to 16 characters; cannot include the period (".") character.
Port Number
The remote port on the server that is to receive the worklist messages.
The default is port 104.
Test
Used to test the connection once you have entered the configuration
data. You can test using either Ping or C-Echo (DICOM Verify). Ping
tests the connection across the network; C-Echo additionally tests that
the DICOM server is active.
For security reasons, some network and network servers are set up
not to respond to Ping or C-Echo commands. Check with your IT
department or PACS Administrator to determine how your network
handles these functions.
NOTE: If you enable Worklist but your DICOM server does not support this feature,
the system will not perform the Worklist function.
Freestyle User Manual
ACUSON
/
/
.
DICOM
WORKLIST
Freestyle. The alphanumeric characters entered in this
ACUSON
Connectivity
8-9