Technical Description - Siemens ACUSON Freestyle User Manual

Diagnostic ultrasound system
Hide thumbs Also See for ACUSON Freestyle:
Table of Contents

Advertisement

Safety and Care

Technical Description

This Section provides a summary of important information regarding the specifications and
safe operation of the
ACUSON
important information on the safe installation and operation of the
particular, the Safety chapter.
Upon request, Siemens Medical Solutions USA, Inc. will make available circuit diagrams,
component part lists, descriptions, calibration instructions, or other information that will assist
service personnel to repair those parts of the system which are designated by Siemens
Medical Solutions USA, Inc. as repairable by non-Siemens personnel.
ACUSON Freestyle Ultrasound System Technical Description
Manufacturer
Electrical Power Requirements
Classifications
Environmental Conditions
Physical Specifications
Batteries
2-48
Freestyle Ultrasound System. Read this entire User's Manual for
Siemens Medical Solutions USA, Inc.
685 Middlefield Road
Mountain View, CA 94043 USA
Input Voltage : 100 – 240 V
Frequency : 50/60 Hz
Current : 1.5 – 0.7 A
Fuse Rating : 250 V, 3 A
Fuse Type: 5 x 20 mm Slo-Blo
The
Freestyle is classified under IEC60601-1, UL 60601-1 as:
ACUSON
1. Class 1 protection against electrical shock. When probe and/or console are
operated from battery power, Class is Internally Powered Equipment.
2. For continuous mode of operation
3. Equipment not suitable for use in the presence of flammable anesthetic
mixture with air or oxygen or nitrous oxide.
4. System: Ordinary equipment without protection against ingress of water.
5. Probe: IPX8
6. Probe : Applied part Type B.
Indoor Use Only
Operating:
Temperature: 10º C to 40º C (50º F to 104º F)
Humidity (Relative Humidity): 10% to 75%
Pressure: 700 hPa to 1060 hPa (20.7 inHg to 31.3 inHg)
Transportation and Storage:
Temperature: –20º C to 60º C (–40º F to 158º F)
Humidity (Relative Humidity): 10% to 90%, Non-condensing
Pressure: 500 hPa to 1060 hPa (14.8 inHg to 31.3 inHg)
Dimensions:
Console: 158 mm L x 373 mm W x 335 H
Probe: 153 mm L x 65 mm W x 34 mm H
Weight: 10.5 pounds (System and one probe with probe battery)
Probe battery reorder number: 11002304
Battery Performance
(Figures approximate and will vary with operating conditions)
Charge Time (empty to full): Probe battery = 2 hours / System battery = 5 hours
Run Time (full to empty): Probe battery = 1.5 hours / System battery = 1 hour
Number of charge cycles before charge capacity reduced to < 70% of original
capacity: Probe battery = 500 / System battery = 300