Connecting the service PC

1.
2.
3.

Configuring and calibrating the tungsten (W) anode

1.
2.
3.

Checking the rotation movement on the new filter disk

1.
2.
3.
NOTICE
MAMMOMAT 3000 Modular
Service
Connect a service PC to the MAMMOMAT 3000 Modular.
Switch on the MAMMOMAT 3000 Modular.
Start the service program.
In "Main menu", choose "Configuration", then "Anode". Set the value to "Enable"
and save with <F2>.
In "Main menu" choose "Configuration" and then "Power". Check that large
tungsten focus is programmed to 4.7 kW and small tungsten focus is
programmed to 0.85 kW. If it is not, change it and save with <F2>.
Check the line resistance and reduce the generator power if necessary (see
installation instructions, measuring the line resistance).
Switch on the MAMMOMAT 3000 Modular.
Mount the magnification table and select W/Rh on the control panel. Check that
there is sufficient space between the cables and the rotating filter assembly.
Select different anode/filter combinations on the control panel. During the
operation of the filter-assembly motor, check that the filter assembly rotates
smoothly without any tendency of jamming. Should this not be the case, this is
probably caused by too tight a fit between the cog disc and the motor pinion. This
can be adjusted by following the instructions below.
Perform step 4-13 only if the result of the functional check
according to point 2 under "Checking the rotation movement on
the new filter disk" above is unsatisfactory.