Chapter 5. Using the Configuration Menu

5.1.1 The Probe Configuration Screen

Tap the Probe Config button to open the Probe Configuration screen (see Figure 53 below).
From the Channel vs. Mode grid, you can activate the installed probes on each channel. See Table 8 on
page 61 for the available measurement modes and units. The probe options vary with the Mode selected, as
follows:
Hygro - M-Series, MISP/MISP2 (MIS), Constant Value (kH) or M-Series with Computer-Enhanced
Response* (M-CER)
Note: Only M-Series PR probes will work with Computer-Enhanced Response. On the main display, the
measurement label will appear in blue.
Temp - M-Series, MISP/MISP2 ( MIS ), Constant Value ( kT )
Pressure - MISP/MISP2 ( MIS) , Auxiliary Input 1 (Aux 1 ), Auxiliary Input 2 ( Aux 2) or Constant Value
( kP )
Oxygen - GE01-GE09, % O2 (Percent), PPM O2 (Parts per Million), PPB O2 (Parts per Billion)
[see "Configuring Delta F Oxygen Sensor Probes" on page 63 for details.]
Aux 1 - mA (current), Volts (voltage)
Aux 2 - mA (current), Volts (voltage)
IMPORTANT: Activating probes that are not physically connected to the meter will result in error messages and
slow performance.
60
Figure 53: Probe Configuration Screen
moisture.IQ User's Manual