GE moisture.iq User Manual page 118

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Chapter 9. Specifications
M Series Probe
Type: Aluminum oxide moisture sensor probe (patented)
Impedance Range: 50 k to 2 M at 77 Hz (depending on vapor pressure of water)
Calibration: Each sensor is individually computer-calibrated against known moisture concentrations,
traceable to national standards.
Operating Temperature: -110 to +70°C (-166 to +158°F)
Storage Temperature: maximum of 70°C (158°F)
Operating Pressure (depends on mount)
M1: 5 microns Hg to 75 psig
M2: 5 microns Hg to 5000 psig
Flow Range
Gases: From static to 10,000 cm/sec linear velocity at 1 atm
Liquids: From static to 10 cm/sec linear velocity at density of 1 g/cc
Built-in Temperature Sensor
Type: Non-linear thermistor
Range: -30°C to +70°C (-22°F to +158°F)
Accuracy: ±0.5°C (± 0.33°F) overall
Response Time: Maximum 1 second in well stirred oil, 10 seconds in still air for a 63% step change in
increasing or decreasing temperature
106
moisture.IQ User's Manual