Chapter 9. Specifications; Electronics - GE moisture.iq User Manual

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Chapter 9. Specifications

9.1 Electronics

Intrinsic Safety
All non-auxiliary inputs are intrinsically safe through internal isolation and energy-limiting circuitry.
Inputs
Two module bays are available. Each module bay can accommodate a 1-channel or 3-channel module. Each
channel can accept:
1 Moisture input (MIS probe or M-Series probe)
1 Temperature input (MIS Probe or M-Series probe)
1 Pressure input (MIS probe)
1 Oxygen input (electrochemical sensor)
2 Auxiliary inputs
Analog Outputs
2 per available channel
High/Low Measurement Alarm Relays
2 per available channel
System Fault Condition Alarm Relay
1 available (not channel specific)
Dimensions (w x h x d, Weights and IP/NEMA Rating
Rack Mount: 482 x 133 x 357 mm (19.0 x 5.2 x 14.1 in.); 11.2 kg (24.7 lb)
Bench Mount: 440 x 133 x 357 mm (17.4 x 5.2 x 14.1 in.); 10.4 kg (22.9 lb)
Panel Mount: 542 x 201 x 357 mm (21.4 x 8.0 x 14.1 in.); 11.3 kg (25.0 lb)
Weatherproof Stainless Steel: 508 x 508 x 229.8 mm (20.0 x 20.0 x 9.05 in.); 24.9 kg (55 lb); IP 66; Type 4X
Weatherproof Fiberglass: 612.5 x 625 x 243 mm (24.11 x 24.61 x 9.57 in.); 24.9 kg (55 lb); Type 4X
Explosion-Proof: 590.6 x 590.6 x 304.8 mm (23.25 x 23.25 x 12.0 in.); 113.6 kg (250 lb); IP 66; Type 4X
Note: See BHGE drawing 712-1889 in the CSS document at the end of this manual.
Power
AC configuration: Universal power supply adjusts automatically from 100 to 240 VAC, 50/60 Hz,
60 Watts maximum, fuse T4A, 250 VAC, 5 x 20 mm
DC configuration: 24 V nominal (± 10%), 60 Watts maximum
moisture.IQ User's Manual
Chapter 9. Specifications
99

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents