Chapter 5. Using the Configuration Menu

5.1.3 Calibrating Individual Channels

1. From the main screen, tap Configuration, then Probe Config, then Reference Values to access the
Channel Calibration screen.
2. In the Channel Calibration screen, enter the high and low values supplied with your dummy probes in
their respective boxes under External References. Tap the Check button on the Channel Calibration
screen after both values have been entered. Once this is completed, these values will be populated for
every channel and do not need to be re-entered for subsequent calibrations.
3. Select the Channel Number for the desired channel to configure. Make sure this is the channel to which
your calibration probes will be attached.
4. Tap the Lab Cal button to begin the low value calibration. The system will check for an active AutoCal
before proceeding. The status should read "Connect probe to channel X, then click Read". Connect your
low value dummy probe to the M-Series input on the corresponding channel of the moisture.IQ module.
5. Tap the Read button. Wait up to a minute. When complete, the status will show "Reading complete on
channel X*."
6. Disconnect the low value dummy probe and connect the high value dummy probe. Do not change the
external reference values or tap the Check button.
7. Tap Lab Cal again, and wait for the message, "Connect the probe to channel X, then click Read." Tap
Read and wait 2-3 minutes.
8. When complete, the new High and Low values will appear in their respective boxes under Channel
Calibration. You can now exit the Channel Calibration screen with the Return button, or repeat steps
3-8 for any other channel requiring calibration.
64
Figure 56: Calibrating Channels
moisture.IQ User's Manual