1.8.3 Adapting an Existing M-Series Probe Cable

If an existing moisture measurement system uses a
adapted for use with an MISP2 probe. To retrofit the existing system, refer to Figure 6 below and complete
the following steps:
1. Locate the existing M-Series probe and disconnect the cable from the probe.
2. Replace the M-Series probe with an MISP2 probe.
3. Using BHGE adapter cable 704-1649-00, connect one end of the cable to the MISP2 probe and connect
the other end to the existing M-Series cable connector.
4. At the moisture analyzer end of the M-Series cable, locate the same M-Series cable and disconnect all
flying leads from the pluggable connector.
5. Locate the GREEN and RED wires of the M-Series cable and connect them to the BLACK and WHITE wires
at the MISP2 probe gray pluggable connector. (The GREEN wire is connected to the MISP connector
"BLK" label and RED wire is connected to the MISP connector " WHT " label.) See "Moisture Image Series
Probes (MISP and MISP2)" on page 19 for help.
6. Trim all other flying leads from the M-Series cable flush with the jacket.
7. Plug the newly-wired MISP2 probe pluggable connector into the moisture.IQ.