Delta F Oxygen Cell
Type: Non-depleting coulometric electrolytic oxygen sensing cell
Available Cells
ppb O
Range: DFOX-1, 0 to 500 ppb/ 5 ppm/ 50 ppm, 1/4 VCR ±3% of reading or 25 ppb
2
ppm O
Range
2
DFOX-9, 0 to 1/10/100 ppm, 1/4 VCR
DFOX-2, 0 to 1/10/100 ppm, 1/8 compression ±3% of reading or 50 ppb
DFOX-3, 0 to 10/100/1000 ppm, 1/8 compression ±3% of reading or 200 ppb
DFOX-4, 0 to 100/1000/10,000 ppm, 1/8 compression ±3% of reading or 2 ppm
DFOX-5, 0 to 50/500/5,000 ppm, 1/8 compression ±3% of reading or 1 ppm
% O
Range
2
DFOX-6, 0 to 5%, 1/8 compression ±3% of reading or 10 ppm
DFOX-7, 0 to 10%, 1/8 compression ±3% of reading or 20 ppm
DFOX-8, 0 to 25%, 1/8 compression ±3% of reading or 50 ppm
Sensitivity: less than 5 ppb (0 to 500 ppbv range)
Response Time: fast response to O
Ambient Temperature: 32 to 120.02°F (0 to 49°C)
Background Gas Compatibility: STAB-EL® cell: All gas compositions including those containing "acid" gases
such as CO
, H
S, C
2
2
Hazardous Area Classification
BASO1ATEX1O98X
II 1G Ex ia IIC T5 Ga
T
= –4°F to +122°F (–20°C to +50°C)
amb
US/CAN Class I, Division 1, Groups A, B, C, D, T4
European Compliance
Complies with EMC Directive 2004/108/EC when connected to a moisture.IQ, Moisture Image
®
Moisture Image
Series 2 or Moisture Monitor™ Series 3 analyzer
moisture.IQ User's Manual
change, equilibrium time is application specific
2
, NOx, SO
, etc.
l2
2
Chapter 9. Specifications
®