GE moisture.iq User Manual page 149

Multi-channel moisture and oxygen analyzer
Hide thumbs Also See for moisture.iq:
Table of Contents

Advertisement

Function Parameter
Channel 6 /
Alarm B
5000
Block
Measurement
Read
5100
Single
Measurement
Read
moisture.IQ User's Manual
Table 13: Modbus Register Map (cont.)
Sub
Modbus
Parameter
ID
Address
Lower
608 3608
Setpoint
Upper
610 3610
Setpoint
Status
620 3620
State
622 3622
Units
624 3624
Type
626 3626
Lower
628 3628
Setpoint
Upper
630 3630
Setpoint
Block Read
0
5000
Measurement
s
Block Read
0
5000
Unit Codes
Block Write
0
5000
Unit Codes
Type
Access
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit Integer RO
Read Input
(0x04)
32-bit Integer RO
Read
Holding
(0x04)
32-bit Integer RW
Read
Holding
(0x03)
32-bit Integer RW
Read
Holding
(0x03)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit float
RW
Read
Holding
(0x03),
Write
Multiple
(0x10)
32-bit float
Read Input
(0x04)
32-bit Integer
Read
Holding
(0x03)
32-bit Integer
Write
Multiple
(0x10)
Appendix D. Modbus Register Map
Modbus
Qty. of
Request
Registers
2
2
2
1 = Energized
0 = De-energized
2
0 = Disabled,
1 = Enabled
2
Measurement/Unit
Code 0 = off,
others see tab
2
1 = Setpoint,
2 = In Band,
3 = Out Band
2
2
2-32
Returns 1-16
measurements
2-32
Returns 1-16 Unit
Codes
2-32
Set Unit Codes for
Measurements 0-15
Notes
137