To Measure Automatically

The Measure button in the menu area activates the automatic measurement function.
The instruction below shows the powerful measurement function of DS1000B series
oscilloscopes.
Auto Measurement
Menu explanation:
Press the Measure button to display the menu of the Automatic Measurements
settings.
The oscilloscopes provide 22 auto measurements: Vpp, Vmax, Vmin, Vtop, Vbase,
Vamp, Vavg, Vrms, Overshoot, Preshoot, Freq, Period, Rise Time, Fall Time,Delay A
Æ B , Delay A Æ B , Phase A Æ B , Phase A Æ B , +Width, -Width, +Duty, -Duty
(10 voltage and 12 timing measurements).
Figure 2-116
Table 2-77 The Measure menu (Page 1/2)
Menu
Source
Voltage
Time
Clear
© 2008 RIGOL Technologies, Inc.
Figure 2-115
The Measure setup button
Settings
CH1, CH2,
Select CH1 or CH2 as source channel
CH3, CH4
for measurement.
Select measure voltage parameter.
Select measure time parameter.
Clear measurement result on screen.