Example 9: Pass/Fail Test - Rigol DS1102/4B User Manual

Ds1000b series digital oscilloscopes
Table of Contents

Advertisement

RIGOL

Example 9: Pass/Fail Test

The Pass/Fail Test is one of enhanced special functions based on DS1000B series. In
this test function the oscilloscope compares the input signal to the established
waveform mask. If the waveform "touches" the mask, a "Fail" signal occurs,
otherwise the test passes. When needed, a programmable output can be used for
external automatic control applications. The output is built in as a standard feature,
and is photo electricity isolated.
Execute the steps as following:
1.
Press Utility
Pass/Fail.
2.
Press Enable Test and select ON.
3.
Press Mask Setting Load.
4.
Press Load to recall the saved mask or press X Mask and Y Mask to adjust the
horizontal limit and vertical limit with the multifunctional knob
Create Mask to create a new mask.
5.
Press Output to select the expected outputting waveforms.
6.
Press Operate to start the test.
3-14
Figure 3-11
Waveform display
User's Guide for DS1000B Series
then press
© 2008 RIGOL Technologies, Inc.

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ds1062/4bDs1202/4b

Table of Contents