Transition Counter
Transition Counter
Used for both AGC and symbol detection routines (see
8-10
and "Minimum Bar Transitions" on page 7-13.
Definition:
During the read cycle, the scanner counts the number of bar and space
transitions that have been defined in the Transition Counter Thresh-
old. If the count matches or exceeds the threshold, the scanner will
perform AGC routines.
Transition Counter Status
Note: This command is currently unavailable in ESP. To determine the current status,
send a
<K505?>
command.
Serial Cmd:
<K505, transition counter
Default:
Disabled
Options:
0 = Disabled
Transition Counter Threshold
Definition:
Determines the number of bar/space transitions that must be read
before a symbol is considered present and AGC can be resolved.
Usage:
Can be useful in certain application where extraneous objects in the
field of view might require a higher threshold.
Serial Cmd:
<K505,
transition counter
Default:
14
Options:
0 to 255
MS-3 Laser Scanner User's Manual
7-12
"No Symbol Message" on page
status,transition counter threshold>
1 = Enabled
status,transition counter
threshold>