Advertisement

5.4.5.10

Measuring point switching (function 59)

Sample selection
M.pt. 1
M.pt. 2
M.pt. 3
M.pt. 4
M.pt. 5
M.pt. 6
Figure 5-44
With this function, you can allocate a maximum of six measuring points and switch them
cyclically and automatically.
The condition is that you parameterized the measuring point relay beforehand using function
71 "Relay outputs", which then actuates the corresponding solenoid valves.
Every measuring point relay is also allocated a time period, which you can enter in the
respective input field. Values between 0 and 60,000 (minutes) are possible for this input.
Press the fifth softkey to activate/deactivate the measuring point switching.
You can allocate a signal relay to each measuring point relay. This allows measuring point
identification separate from the measuring point relay. Configure the signal relay, also using
function 71.
See also
Relay outputs (function 71) (Page 82)
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
Rel. 2 :
45: min
Rel. 1
2O
Rel. 4
12
Rel. 3
8
Rel. 6
23
Rel. 5
6
M.p. switching on/off
Measuring point switching
O
2
min
min
min
min
min
Operation
5.4 Operating functions
79