Electrical Connections - ABB EasyLine EL3000 Series Datasheet

Continuous gas analyzers
Hide thumbs Also See for EasyLine EL3000 Series:

Advertisement

Electrical connections

Power supply and signal lines model EL3020
X01
1
Power supply and signal lines model EL3040
M25
M32
M25
M25
M20
M32
1
Power supply connection
(3-pin plug per EN 60320-1/ C14; connection cable
supplied)
I/O modules (4 slots), options:
2
Digital I/O module (max. 3 modules)
3
Analog output module (max. 2 modules)
4
Modbus module (RS232 and RS485 interface)
5
Profibus module (RS485 and MBP interface)
6
Ethernet-10 /100BASE-T interface (8-pin RJ45 plug)
7
Potential compensation connection (max. 4 mm
Screwed cable glands for cable diameter:
M20 Power supply
M25 Modbus / Profibus
M25 Network
M25 Analog outputs
M32 Digital inputs / outputs
M32 Digital inputs / outputs
10/24-4. 1 0 EN Rev. 9
(view from behind)
2
-X26
-X27
3
-X24
-X25
4
-X22
-X23
5
-X20
-X21
6
7
10/100
BASE-T
(view from below)
2
-X26
3
-X24
4
-X22
5
-X20
10/100
6
BASE-T
1
2
)
5–13 mm
8–17 mm
8–17 mm
8–17 mm
12–21 mm
12–21 mm
EasyLine EL3000 Series Data Sheet
-X27
-X25
-X23
-X21
7
Notes
Both drawings show examples for the I/O modules equipment.
I/O module connection:
– The maximum capacity of terminals for stranded or solid
conductors is 1 mm
– The stranded conductor may be tinned on the tip or twisted
for simplified connection.
– When using wire end ferrules the total section should not
2
exceed 1 mm
, i.e. the maximum stranded conductor section
2
is 0.5 mm
. The Weidmüller PZ 6/5 crimping tool must be
used for crimping the ferrules.
Functional scope of the Ethernet interface:
– Communication with configuration software ECT for gas
analyzer configuration and software update.
– QAL3 data transfer if the QAL3 monitoring option is integrated
in the gas analyzer.
2
(17 AWG).
17