OPErATiON
PATiENT dATA CArd- SET uP OF NEw CArd (CONTiNuEd)
Access Electrode Placement
The following information uses an IFC
Treatment as an example . Electrode
Placement procedures for all modalities are
performed in the same basic fashion .
Press the Electrode Placement button .
ElEcTrODE PlacEMEnT buTTOn
Electrode Placement Set Up
Press the Pad button to select the Red (+) or
Black (-) electrode .
Press the Chan button to select Channel 1
or Channel 2 .
Press the Side button to select Front, Back,
Left, or Right of the body graphic .
Press the Size button until desired electrode
size is displayed . If electrode desired is not
listed, select Other .
PaD
buTTOn
chan
buTTOn
SIDE
buTTOn
SIzE
buTTOn
NOTE: When Ultrasound is the modality,
only the Side button is available .
82
Vectra Genisys® Therapy System
Electrode Placement
Press the Up, Down, Left and Right Arrow
buttons to position the selected electrode
as close to the actual treatment location as
possible .
Press the Pad button to select the other
electrode . Repeat above procedure for
electrode positioning .
If applicable, press the Chan button to
select another channel and repeat above
procedures .
After positioning the electrodes, press the
Accept and Return Arrow button .
accEPT anD rETurn
arrOw buTTOn
ElEcTrODES
POSITIOnED