PATiENT PrEPArATiON
uLTrASOuNd PATiENT PrEPArATiON
Preparing Treatment Area
Examine the skin for any wounds and clean the skin .
Size of Applicator
View the Sound Head Recommendation in the Treatment
Review screen for Ultrasound (as a reference point only) prior to
administering treatment .
Sound Heads are available in the sizes shown below .
2
2
1 cM
2 cM
Applicator Preparation
Clean applicator before each therapy session with warm soapy
water .
2
5 cM
2
10 cM
(STanDarD)
Conductive Medium
Liberally apply Conductor™
Transmission Gel or equivalent to
the treatment area on the patient .
Treatment Area
Move the Sound Head during
therapy session in a circular
motion . The area treated should
be two times the diameter of the
Sound Head .
Applicator Coupling
If US Coupling is On, the Sound
Head is properly coupled to
the patient and administering
ultrasound when the LED is
constantly illuminated .
NOTE: Refer to
page 29
for US
Coupling settings .
36
Vectra Genisys® Therapy System
aPPly
cOnDucTIvE GEl
cOuPlInG
InDIcaTOr
lED