Removal- Channel 3/4 Electrotherapy, Nimh Battery, And Laser Module; Disconnect Mains Power; Remove Lead Wires And Accessories; Remove Therapy System From Cart - Vectra Fitness Genisys 2761 User Manual

Two/four channel electrotherapy / combination system / with cart, no semg, therapy system cart assembled / unassembled, nimh battery module, dual channel semg module, user remote controls, channel 3/4 electrotherapy module
Table of Contents

Advertisement

iNSTALLATiON/rEMOvAL

rEMOvAL- ChANNEL 3/4 ELECTrOThErAPY, NiMh BATTErY, ANd LASEr MOduLE

Disconnect Mains Power

Disconnect the system from the power
source (outlet or remove battery module
if installed) before attempting any
maintenance, installation, removal, or
replacement procedures to prevent electrical
shock and possible damage to system .
Disconnect the Mains Power Cord from
the power supply . Remove the Rear Access
Panel and disconnect the Mains Power
Cord from the Therapy System .
MaInS
POwEr
cOrD

Remove Lead Wires and Accessories

Remove the Front Access Cover and
disconnect the Lead Wires and Accessories
from the Therapy System and Channel 3/4
Electrotherapy Module .
rEMOvE lEaD
wIrES anD
accESSOrIES
rEar
accESS
PanEl
NOTE: Keep Lead Wires and accessories
for later re-installation to the
Therapy System .

Remove Therapy System from Cart

Remove the Therapy System from the
Therapy System Cart, if equipped . Refer
to
instructions .
Place Therapy System face down on a clean
working surface protected with a soft, clean
fabric to prevent damage to the lens .
NOTE: It is not necessary to remove
123
Vectra Genisys® Therapy System
pages 131 through 133
for proper
the sEMG Module from the Channel
3/4 Electrotherapy Module, if installed .

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents