iNSTALLATiON/rEMOvAL
iNSTALLATiON- ChANNEL 3/4 ELECTrOThErAPY, NiMh BATTErY ANd LASEr MOduLE (CONTiNuEd)
Disconnect Mains Power
Disconnect the system from the
power source (outlet or remove battery
module if installed) before attempting
any maintenance, installation, removal or
replacement procedures to prevent electrical
shock and possible damage to system .
Disconnect the Mains Power Cord from
the power supply . Remove the Rear Access
Panel and disconnect the Mains Power
Cord from the Therapy System .
MaInS
POwEr
cOrD
Remove Lead Wires and Accessories
Remove the Front Access Cover and
disconnect the Lead Wires and Accessories
from the Therapy System .
rEMOvE lEaD
wIrES anD
accESSOrIES
rEar
accESS
PanEl
Remove Therapy System from Cart
Remove the Therapy System from the
Therapy System Cart, if equipped . Refer
to
instructions .
Place Therapy System face down on a clean
working surface protected with a soft, clean
fabric to prevent damage to the lens .
NOTE: Do not remove the sEMG
118
Vectra Genisys® Therapy System
pages 131 through 133
for proper
lanyarD
Module, if installed . The
sEMG Module will not
interfere with installation
of the Channel 3/4 Electrotherapy
Module .
EMG
s
MODulE