iNSTALLATiON/rEMOvAL
iNSTALLATiON- ChANNEL 3/4 ELECTrOThErAPY, NiMh BATTErY ANd LASEr MOduLE (CONTiNuEd)
Specifications
DEPTh
NOTE: All waveforms except High Voltage Pulsed Current (HVPC)
of the Vectra Genisys Therapy System have been designed
with a 200 mA current limit .
VMS™, VMS™ Burst and all TENS waveform output intensities are
measured, specified and listed to peak, not peak to peak .
Dimensions
Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 . 2 50 in (21 cm)
Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11 . 8 75 in (30 cm)
Height . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 . 5 00 in (11 . 5 cm)
Weight
Standard Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . 0 lbs ( . 5 0 kg)
HEIGHT
Power
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .System Dependent
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .System Dependent
Electrical Class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CLASS I
Electrical Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .TYPE BF
WIDTH
Regulatory Compliance
UL/IEC/EN 60601-1
IEC 60601-2-10
Waveform & Current Specifications
All waveform/currents available to the Vectra Genisys Therapy
System are available to the Channel 3/4 Electrotherapy Module
once installation is complete . Refer to
available waveform specifications .
117
Vectra Genisys® Therapy System
pages 19 through 23
for