SPECiFiCATiONS
wAvEFOrM SPECiFiCATiONS (CONTiNuEd)
Premodulated (Traditional 2 Pole IFC)
Premodulated Current is a medium frequency waveform . Current
comes out of one channel (two electrodes) . The current intensity
is modulated: it increases and decreases at a regular frequency
(the Amplitude Modulation Frequency) .
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Electrodes
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-100 mA (CC)
Carrier Frequency (Fixed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2,500 Hz
Beat Frequency (Sweep Off ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-200 Hz
Sweep Time (Fixed) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 seconds
Sweep Low Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-199 Hz
Sweep High Beat Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-200 Hz
Vector Scan . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Off, Manual, 40%, and 100%
Mode Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CC or CV*
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Min
Available on Channel . . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option
*CC= Constant Current
CV= Constant Voltage

High Voltage Pulsed Current (HVPC)

The High Voltage Pulsed Current (HVPC) has a very brief pulse
duration characterized by two distinct peaks delivered at
high voltage . The waveform is monophasic (current flows in
one direction only) . The high voltage causes a decreased skin
resistance making the current comfortable and easy to tolerate .
Output Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrodes or Probe
0- 100 V (CV)
Output Intensity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0-500 V
Polarity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Positive or Negative
Ramp . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 . 5 sec, 1 sec, 2 sec, 5 sec
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Peak Current or Volts
Sweep . . . . . . . . . . . . . .Continuous, 80/120 pps, 1/120 pps, 1/10 pps
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-120 pps
Cycle Time . . . . . .Continuous, 5/5, 4/12, 10/10, 10/20, 10/30, 10/50
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-60 Min
Anti-Fatique . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Off or On
Available on Channels . . . . . . . . . . . . . . . . . . . . . . . . 1 & 2, 3 & 4 Option