Dräger Babylog 8000 plus Operating Instructions Manual page 85

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Analog and Digital Interface
(Available Option)
Interface provides analog output of measured values,
output of data reports, communication with patient
monitors or personal computers, e.g. with EvitaView
PC software, allowing graphic and numeric display of
ventilation parameters.
WARNING !
Therapeutical decisions should not be made
solely on the basis of the data transmitted via
the communications interface. Always use other
diagnostic means in addition.
Place connected devices in the same room with the
Babylog 8000 plus ventilator, but at a distance of at
least 1.5 m (5 ft) from the patient.*
The two analog outputs supply each one of the
available measured values.
The RS 232 serial interface outputs information to a
printer:
-
reports
-
trend information
-
waveform data
or transmits data to a patient monitor or a personal
computer.
The marker output
events:
-
triggered mandatory breath
-
mandatory breath
-
alarm.
Analog Data Output
The two analog outputs Analog 1 and Analog 2 each
provide one measured value, e.g. V
Voltage range: 0 to 10 V.
Connect recorder (impedance ≥ 1 MΩ) with interface
cable 83 06 487 (see also Technical Data, page
127).
Select signal and range, see page 92.
___________
* Requirement pursuant EN 60601-1-1
indicates certain
, P
, Flow ... .
T
aw
Analog and Digital Interface
Analog1
Analog2
Operation
85