Dräger Babylog 8000 plus Operating Instructions Manual page 120

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Technical Data
Environmental Conditions
Settings - Range, Resolution, and Accuracy
Note: Specifications are given conforming to ISO 5369.
Environmental Conditions
During operation:
Temperature
Atmospheric pressure
Rel. humidity
For storage:
Temperature
Atmospheric pressure
Rel. humidity
Settings - Range, Resolution, and Accuracy
Parameter
Inspiratory time T
in
Expiratory time T
ex
O
Conc.%
2
Inspiratory flow *
insp
Expiratory flow *
exp
P
insp
PEEP
alarm limits
*
E
Delay for lower alarm limit
*
E
Apnea time
Tachypnea
Tidal volume V
TSet
120
10 to 40 °C
780 to 1060 hPa
30 to 90 % (no condensation)
–20 to 60 °C
500 to 1060 hPa
10 to 95 % (no condensation)
Range
0.1 to 2 s
0.2 to 30 s
21 to 100 %
1 to 30 L/min
1 to 30 L/min
10 to 80 cmH
O
2
0 to 25 cmH
O
2
0 to 15 L/min
0 to 30 s
5 to 20 s
20 to 200 bpm
2 to 100 mL
Resolution
0.1 s to 1 s: 0.01 s
1 s to 2 s: 0.1 s
0.2 s to 1 s: 0.01 s
1 s to 10 s: 0.1 s
10 s to 30 s: 1 s
1 %
1 to 10 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
1 to 10 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
1 cmH
O
2
to 10 cmH
O: 0.1 cmH
O
2
2
>10 cmH
O: 1 cmH
O
2
2
1<1 L/min: 0.01 L/min
1 to 10 L/min: 0.1 L/min
1 s
1 s
5 bpm
2 to 9.9 mL: 0.1 mL
10 to 19.5 mL: 0.5 mL
20 to 100 mL:
1 mL
Accuracy
±10 ms
±10 ms
±3 %
±10 %
±10 %
O ±3 %
1 cmH
2
(measuring function)
O ±3 %
1 cmH
2
(measuring function)
±10 % (measuring function)
±10 ms
±10 ms
1 bpm
up to 5 mL: ±0.5 mL
> 5 mL: ±10 %