Pressure measurement
out of range
Printing cancelled
Check printer!
Re-calibrate flow sensor
if exchanged!
V
low
T
Check settings!
Tube obstructed?
Ventilator malfunction
Error code: xxx
Cause
Liquid in patient block.
Condensation in patient circuit.
Fault in pressure measurement
function.
ID of patient circuit too small
Printer switched off.
No paper.
Faulty cable.
RS232 interface of ventilator or printer
not properly configured.
During operation:
Flow sensor has been exchanged
Cable was disconnected and
re-connected.
Set tidal volume V
has not been
Tset
delivered.
Kinked or blocked tube.
Auto-triggering
Ventilator faulty.
Remedy
Replace patient block
Remove condensation.
Take ventilator out of service.
Use circuit with less resistance.
Switch on printer.
Load paper.
Replace cable.
Configure RS232 interface,
see page 93, 94.
Press »Confirm« key.
Calibrate flow sensor, page 40.
.
Press »Confirm« key.
Increase inspiratory flow *
increase inspiratory time,
increase P
if necessary.
insp
Clear passage.
Correct trigger sensitivity.
Take ventilator out of service.
Troubleshooting
,
insp
109