Calibrating the O
-Sensor Manually
2
After sensor has been replaced, but possible at all
times.
1 Press »Cal. Config.« menu key.
2 Press »O
-Cal« menu key.
2
3 After approximately 5 minutes, O
from the display, calibration is now completed.
To clear the respective text message:
1 Press »Confirm« key.
cal disappears
2
1