Preparation
Installing an O2 Sensor Capsule
Installing an O
Sensor Capsule
2
– before first use
– when calibration can no longer be performed.
Unscrew the two screws in cover on the right and
remove cover lid.
Pull used O
sensor capsule.
2
Insert new capsule into opening - circular contacts
must point towards cover lid.
Press down cover and re-tighten the two screws.
After installation:
Wait for 15 minutes for sensor to warm up before
calibration.
Calibrate sensor manually, see page 39.
Dispose of the used sensor, see page 104.
WARNING !
Treatment of batteries and O2-sensor
capsules:
Do not throw into fire! Risk of explosion.
Do not force open! Danger of bodily injury.
Follow all local, state, and federal regulations
with respect to environmental protection
when disposing of batteries and O2-sensor
capsules.