Care
Disassembly
Disassembly
Clean and process ventilator after each patient.
WARNING !
Always follow accepted hospital procedures
for handling equipment contaminated with
body fluids.
Patient breathing circuit
Note: The following description refers to a disposable
circuit system. Refer to humidifier operating instructions
for disassembly of humidifier components.
Empty water trap in expiratory limb, if necessary.
Unplug connector of flow sensor at the wye and at the
back of ventilator.
Remove breathing circuits from flexible circuit support
arm in the incubator or from hinged arm.
Detach circuit from ports on ventilator.
Detach flow senor.
Detach temperature sensor(s) from circuit, disconnect
cable from humidifier unit.
Remove sensor element from flow sensor:
Press buttons on both sides of flow sensor element
and gently remove insert from wye.
Removing the Expiratory Valve
1 Push lever upward (release position).
2 Pull expiration valve out (forward).