Dräger Babylog 8000 plus Operating Instructions Manual page 87

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Press »Cal. Config.« key.
Press »Print« menu key
Display (example):
Printing Reports
For documenting measured, set, and status values.
Example:
For printing a single report:
Press »Select« menu key, until »Report« is
highlighted.
Press »Start« button. The function of the button
changes to »Stop«.
NOTE: While a report is being printed, the functions
»Print all« and »BabyLink« are not available. However,
any of the other functions can be started. Babylog will
execute them after finishing the report.
To abort printing:
Press »Stop« button.
+--------------------------------------------------------------------------------+
|7:51
2.08.1997
|
|Settings:
|
|TI:
0,82 s
Pinsp:
33,0 cmH2O
|TE:
1,59 s
PEEP:
6,1 cmH2O
|I:E: 1:
2,0
FiO2:
21 %
|fset: 24,9 bpm
|
Flow in:
8,4 L/min
|
Flow ex:
8,4 L/min
+---------------------------------------+----------------------------------------+
|VG:
|VTset: --,- mL
|
+---------------------------------------+----------------------------------------+
|VIVE:
off
Trigger:
1,3
|
+--------------------------------------------------------------------------------+
Operation
Analog and Digital Interface
CMV
Babylog 8000 plus
|
Measurements:
|
|
FiO2:
21 %
Mean:
15,0 cmH2O|
|
Leak:
0 %
PIP :
33,5 cmH2O|
|
spont:
0 %
PEEP:
6,0 cmH2O|
|
f:
25,0 bpm
|
VT:
24 mL
|
MV:
0,58 L/min
TIspont:
-,-- s
|
LFT:
|
C:
1,22 mL/mbar
R: 408 cmH2O/L/s|
|
C20/C: 3,37
TC: 499 ms
|
Flow measurement:
|
NTPD
Y
FiO2-compensated
87
|
|
|
|
|
|
|
|
|
|
|