Preparation
Checks of Readiness for Operation
Testing Alarm Limits
Apnea alarm limits
1 Switch on CPAP, please refer to page 64.
Make sure that apnea alarm is enabled, please refer
to page 68.
After 30 seconds (max.):
2 Red alarm will start blinking, audible alarm will sound.
Display (example):
Apnea
or
MV low
Airway pressure alarm limits
3 Switch on CMV, please refer to page 58.
4 Block expiratory limb of the circuit (pinch).
2 Red alarm will start blinking, audible alarm will sound.
Display (example):
Hose kinked ?
or
Airway pressure high
Inspiration cancelled
46
2
1
3
4