Dräger Babylog 8000 plus Operating Instructions Manual page 38

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Preparation
Before Using for the First Time
Calibrations
1 Press »Vent. Mode« menu key.
2 Press »CMV« menu key.
Press »On« menu key.
Press »
« key.
Display (example):
3 The alarm limit symbol
request to set alarm limits.
Calibrations
Calibration of the built-in oxygen analyzer:
-
is performed automatically every 24 hours.
-
must be performed manually after replacing a used
O
sensor.
2
-
may be performed manually at any time.
Calibration of the flow sensor:
-
is required every time the ventilator is switched on,
-
after exchanging a flow sensor.
Calibration of pressure sensors:
-
is automatic when ventilator is switched on.
38
will be flashing as a
2
1
3