Dräger Babylog 8000 plus Operating Instructions Manual page 45

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Testing CMV
Switch on CMV, please refer to page 58.
Adjust dial knobs whose green LEDs are lit:
1 »O
-Conc. %« to 21.
2
2 »Insp. Flow * * * * « to 10.
3 »T
« to 0.4.
in
»T
« to 0.6.
ex
4 »P
« to 20.
insp
5 »PEEP/CPAP« to 0.
From the monitoring menu, press menu keys
»PEEP/CPAP« and »Pressure«.
Display (example):
Values displayed should be in acceptable range:
PIP
18 to 22 cmH
P
6 to 10 cmH
2
PEEP
–1.5 to 1.5 cmH
Testing PEEP
6 Set knob for »PEEP/CPAP« to 10.
Confirm setting:
7 Press »Confirm« key.
Display (example):
Values displayed should be in acceptable range:
PIP
18 to 22 cmH
P
12 to 16 cmH
2
PEEP
8 to 12 cmH
2
O
2
O
O
2
O
2
O
O
Checks of Readiness for Operation