Dräger Babylog 8000 plus Operating Instructions Manual page 82

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Operation
Messages
In the Event of an Alarm
To mute message and audible alarm for 30 seconds:
1 Press »Confirm« key.
Each message is recorded automatically in the
ventilator message log.
Silencing audible alarm for 2 minutes
g
2 Press »
« key.
NOTE: New messages are displayed without their
associated audible alarm tone during the 2-minute
silencing period.
To re-arm audible alarm:
g
2 Press »
« key again.
WARNING !
The alarm silence button is intended to provide a
way of muting audible alarms while corrective
action is taken. The operator of the ventilator
must still assume responsibility for proper venti-
lator function and patient safety in the event of
an alarm. Failure to identify and correct alarm
situations may result in patient injury.
NOTE: A troubleshooting list can be found on pages 106
to 109.
Message Log
Every warning, caution, and advisory message is
registered in a message log. Log entries include their
time of occurrence, text of the message, and an
indication of whether the message had been
acknowledged (normal display of message in log) or not
(display of message highlighted).
From the monitoring menu, press »
= Log.
The »
« and »
forward and backward within the log.
Display (example):
To leave menu:
Press »
« menu key.
82
« menu key
« menu keys allow to scroll
2
1
D
Babylog 8000 plus