Dräger Babylog 8000 plus Operating Instructions Manual page 81

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Messages
Messages are displayed according to their urgency in
a hierarchical sequence. If two alarm conditions are
detected simultaneously, the more urgent one will be
communicated first.
Messages appear in a overlay "window" on top of the
respective screen image.
At the same time, one of three specific sound
patterns will be used to communicate the appropriate
level of urgency.
Display (example):
A message will disappear when the causing condition
no longer exists.
Alarm Categories
Warning (top priority message)
blinking red alarm light, intermittent audible alarm.
Warning messages require immediate action to
prevent a life threatening condition, e.g. when a
circuit is kinked.
Caution
blinking red alarm LED, repetitive three tone
sequence.
Caution messages require short term action to
prevent development of a condition with the possibility
of patient injury. Respond appropriately and complete
the action within 2 minutes at the most.
WARNING !
Warning or Caution level audible alarms
require immediate operator attention to avert
or to prevent development of situations with
the possibility of patient injury.
Advisory
individual three tone sequence.
Advisory messages are reminders for special
functions or may highlight a particular setting.
D
Babylog 8000 plus
Operation
Messages
81

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents