Independent Expiratory Flow VIVE
VIVE (Variable Inspiratory and Variable
Expiratory Flow)
Expiratory continuous flow *
independently from inspiratory flow *
mandatory breaths, inspiratory flow is applied. During
spontaneous breathing and during CPAP, the
expiratory flow is in effect.
An increased inspiratory flow *
-
to supply a higher flow for spontaneous breathing
than the one scheduled for ventilator breaths.
-
to improve gas washout in the wye by increasing
turbulence in the circuit system.
-
to allow for separate shaping of manually triggered
breath in CPAP.
A reduced expiratory flow
conserve oxygen and, thereby, reduce cost of
ventilation.
1 Press »Vent. Options« key
Press »VIVE« menu key and use menu keys »
»
« for setting expiratory flow.
Press »On« menu key and return with »
may be set
exp
. During
insp
may be used
insp
*
may be used to
exp
« or
«.