Dräger Babylog 8000 plus Operating Instructions Manual page 103

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

In Case of Ventilator Malfunction
WARNING !
Never operate the ventilator if it has suffered
physical damage or does not seem to operate
properly. In this case always refer servicing to
properly trained and factory authorized service
personnel.
For ventilators with liquid crystal displays (LCD):
WARNING !
If, in case of damage to the ventilator, the
glass of the LCD screen is broken, a liquid
chemical may escape that should not come
into contact with the skin. In case of
contamination, wash off with soap
immediately.
User-Replaceable Parts
Replacing cooling air intake filter
– Clean or replace after 4 weeks.
Replace after 1 year at the latest.
Remove air intake filter from its holder in the back of
the ventilator.
Replace or clean in warm water with detergent
added. Dry well.
Re-insert cooling-air filter into holder.
Verify that it is fit properly.
Dispose of used cooling-air filter with domestic waste.
Care
In Case of Ventilator Malfunction
User Replaceable Parts
103