Dräger Babylog 8000 plus Operating Instructions Manual page 59

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Displaying the flow waveform:
Press »Graph« and »Flow« menu keys.
To leave this menu:
Press »
« key.
Displaying measured pressure values:
Press »Meas« and »Paw« menu keys.
Set the desired ventilation pattern using dial knobs
»P
«, »PEEP«, »T
«, »T
« and »
* * * *
«.
insp
in
ex
insp
Displaying the measured values for volume:
Press »Vol« menu key.
Set dial knobs »P
« and »PEEP« in such a way
insp
that the required tidal volume V
is delivered.
T
For setting of alarm limits for minute ventilation,
please refer to page 70, 71.
Ventilation without a pressure plateau
This represents volume controlled ventilation.
Peak inspiratory pressure results from settings of T
in
and
*
insp.
Tidal volume V
is approximately:
T
C
rs
V
= T
·
·
*
T
in
insp
C
+ C
rs
c
T
= inspiratory time,
in
= continuous inspiratory flow,
*
insp
C
= compliance of patient's respiratory system,
rs
C
= compliance of patient circuit system.
c
Tidal volume is controlled via flow and inspiratory
time.
Operation
Setting Ventilation Modes and Extensions
Page 59