Important Safety Information

Definitions

Summary of WARNINGS and CAUTIONS
Definitions
WARNING !
A WARNING statement refers to conditions
with a possibility of personal injury if
disregarded.
CAUTION !
A CAUTION statement designates the possibility
of damage to equipment if disregarded.
NOTE: A NOTE provides additional information
intended to avoid inconveniences during operation.
Inspection
= examination of actual condition
Service
= measures to maintain specified
condition
Repair
= measures to restore specified
condition
Maintenance = inspection, service, and repair,
where necessary
Preventive
= Maintenance measures at regular
Maintenance
intervals
Typing conventions in this manual
Controls (knobs, hard keys and menu keys) are
designated as »Control Name«, e.g.:
»PEEP/CPAP«
On-screen messages are printed in bold, e.g.:
Calibrate flow sensor!
10
Summary of WARNINGS and
CAUTIONS
General Precautions
WARNING !
Strictly follow this Operator's Instruction
Manual
Any use of the product requires full under-
standing and strict observation of all portions
of these instructions. The equipment is only to
be used for the purpose specified under
"Intended Use" (page 18) and in conjunction
with appropriate airway monitoring (see page
19). Observe all WARNINGS and CAUTIONS
as rendered throughout this manual and on
labels on the equipment.
WARNING !
Babylog 8000 plus is solely intended for use by
physicians or by allied health care profession-
als on the order of a physician. All users must
be properly trained and completely familiar with
these Operating Instructions.
WARNING !
This device is to be used only in rooms with
line power installations complying with
national safety standards for hospital patient
rooms. (e.g., IEC 601.1, "Safety of Medical
Equipment).
To maintain grounding integrity, connect only
to a"hospital grade" receptacle.
Always disconnect supply before servicing.
WARNING !
DANGER, risk of explosion if used in the
presence of flammable anesthetics.
This device is neither approved nor certified
for use in areas where combustible or explo-
sive gas mixtures with air or with nitrous oxide
are likely.