Release circuit again. Remove ET-tube connector
from wye:
Red alarm will start blinking, audible alarm will sound.
Display (example):
Airway pressure low
or
Leak in hose system?
Check setting !
Reset knob for »PEEP/CPAP« to 0.
Reconnect test lung.
WARNING !
The ventilator is ready for operation only when:
- it is completely assembled with all required
auxiliary equipment in place,
- all sensors are calibrated (O
- all checks have been completed successfully.
Check humidifier according to its operating
instructions.
, Flow),
2
Preparation
Checks of Readiness for Operation