To select default settings during operation:
Press »* * * * /P« menu key.
NOTE: If a measured value exceeds the set range,
the output voltage will be clipped.
NOTE: The setting will be retained after the ventilator
is switched off.
Configuring the marker output
Press »Param« menu key repeatedly, until »
highlighted.
Select desired signal for analog output with menu
keys »
« or »
triggered mandatory breath
mandatory breath
alarm condition.
Default setting: mandatory breath.
NOTE: The setting will be retained after the ventilator
is switched off.
Configuring RS232 interface
To select data transmission rate (baud rate) and parity
check mode.
Press »Param« menu key repeatedly, until
»Baud rate« is highlighted.
Select the baud rate with menu keys
»
« or »
«:
9600, 2400, or 1200
Select »Parity» using »Param« menu key.
« is
«:
or
or
Operation
Analog and Digital Interface
93