Parameter
V
T
MV
Leak
DCO
2
Spontaneous MV
Flow waveform
Breathing rate
RVR
Inspiratory oxygen concentration
Sensor
Measuring range
Measuring acuracy
The measured value is displayed numerically on screen.
Parameter
FiO
2
Compliance C of respiratory system
Resistance R
Time constant TC of respiratory
system
Overdistension index C
/C
20
Range
0 to 999 mL
0 to 30 L/min
0 to 100 %
0 to 999 mL
2
/s
0 to 100 %
–20 to 20 L/min
0 to 9.9 /bpm
10 to 999 /bpm
0 to 1000 bpm/mL
Fuel cell inside ventilator exposed to inspiratory flow
18 to 100 Vol.%
±3 Vol.% of pressurized gases O
When calibrating, Babylog 8000 plus uses 20.9 Vol.% O
100 Vol.% O
respectively, for calculation.
2
Range
18 to 100 Vol.%
0.2 to 10 mL/cmH
O
2
10 to 10000 cmH
O/L/s
2
0.1 to 5 s
0 to 5
Technical Data
Measuring Functions - Range and Accuracy
Resolution
0 to 9.9 mL: 0.1 mL
10 to 99 mL: 1 mL
0 to 0.99 L/min: 0.01 L/min
1 to 9.9 L/min: 0.1 L/min
10 to 30 L/min: 1 L/min
0 to 10 %: 10 %
10 to 100 %: 1 %
1 mL
2
/s
1 %
–2.5 to 2.5 L/min: 0.1 L/min
–5 to 5 L/min: 0.2 L/min
–10 to 10 L/min: 0.4 L/min
–20 to 20 L/min: 0.8 L/min
0.1 bpm
1 bpm
0.1 bpm/mL
and Air
2
or
2
Resolution
1 Vol.%
0.1 mL/cmH
O
2
1 cmH
O/L/s
2
0.01 s
0.1
123