Dräger Babytherm 8000 WB Operating Instructions Manual

Dräger Babytherm 8000 WB Operating Instructions Manual

Heated bassinet
Hide thumbs Also See for Babytherm 8000 WB:

Advertisement

Quick Links

®
Babytherm
8000 WB
Heated Bassinet
Operating Instructions

Advertisement

Table of Contents
loading

Summary of Contents for Dräger Babytherm 8000 WB

  • Page 1 ® Babytherm 8000 WB Heated Bassinet Operating Instructions...
  • Page 2 NOTICE Proprietary Information This document contains information in which Dräger, Inc. claimed proprietary rights. The information may not be reproduced in whole or in part except as authorized in writing by Dräger. This information is the property of Dräger, Inc., it is provided solely for the use intended.
  • Page 3: Table Of Contents

    Contents Contents Page Page Introduction READ THIS FIRST......4 Care ..............30 Operator's Responsibility for Patient Safety..4 Disassembly............ 30 Limitation of Liability.......... 4 Cleaning/Disinfecting/Sterilizing.......31 Warranty............5 Before Next Use ..........31 Definitions............6 Inspection and Maintenance....... 32 General WARNINGS and CAUTIONS....7 Replacement of Parts........
  • Page 4: Introduction Read This First

    Introduction Operator's Responsibility for Patient Safety Limitation of Liability Introduction Limitation of Liability Dräger, Inc.'s liability, whether arising out of or related to Operator's Responsibility for Patient Safety manufacture and sale of the goods, their installation, For correct and effective use of the product and in demonstration, sales representation, use, performance, order to avoid hazards it is mandatory to carefully or otherwise, including any liability based upon Dräger,...
  • Page 5: Warranty

    Warranty Introduction Warranty All Dräger products are guaranteed to be free of defects Application of this warranty is subject to the following for a period of one year from date of delivery. conditions: The following are exceptions to this warranty: 1.
  • Page 6: Definitions

    Introduction Definitions Definitions WARNING ! A WARNING statement refers to conditions with a possibility of personal injury if disregarded. CAUTION ! A CAUTION statement designates the possibility of damage to equipment if disregarded. NOTE: A NOTE provides additional information intended to avoid inconveniences during operation. Inspection = examination of actual condition Service...
  • Page 7: General Warnings And Cautions

    We recommend a "hospital grade" receptacle. Always disconnect supply before servicing. obtaining a service contract through your vendor. Maintenance or repair of the Babytherm 8000 WB shall be performed only by Dräger authorized WARNING ! technical service representatives.
  • Page 8: Precautions During Preparation

    The Dräger gel mattress (part no. 2M 20 827) must not be used in any bassinets other than the Babytherm 8000 WB. Heater plate temperatures above 40 °C (104 °F) implicate the danger of patient burns! Use of a gel mattress with unheated bassinets will result in patient cooling.
  • Page 9 . use with the Babytherm 8000 WB. Specifically, combinations with radiant warmers that are not intended for use with heated bassinets bear...
  • Page 10: Precautions During Care

    Preventive Maintenance work on the Babytherm 8000 WB may be performed by trained and factory authorized staff only. CAUTION ! Do not expose Babytherm 8000 WB to high level of ultraviolet radiation. It may cause cracks in the WARNING ! PMMA (perspex) parts.
  • Page 11: Intended Use

    Description Intended Use Description Mattress heating Babytherm 8000 WB is a heated bassinet intended as a The Babytherm 8000 WB features a mattress heating therapeutic device providing warmth for premature and system that uses conductive heat transfer to warm up full-term neonates and infants with a weight up to 16 lbs.
  • Page 12: Preparation

    Adjusting the Height Storage of Supplies Attaching Infusion Bottle Holder Preparation Adjusting the Height (Option) Before operating the Babytherm 8000 WB, always check electric height adjustment: Connect power supply. WARNING ! To maintain grounding integrity, connect only to a "hospital grade" receptacle.
  • Page 13: Installing Surrounding Panels

    Installing Surrounding Panels Preparation Installing Inside Panels Installing Surrounding Panels The bassinet is surrounded by two side panels and two end panels, which can be opened independently of each other. End panels have cutouts with U-grommets for routing circuits and patient cables. Insert lower pins on panel hinges into the left- and right-hand slots in bassinet base.
  • Page 14: Fitting A Mattress

    Preparation Fitting a Mattress Installing Cover Fitting a Mattress Select mattress type: – Operation with mattress heating: Gel mattress 2 M 20 827. – Operation without mattress heating: Foam rubber mattress 2 M 21 012. WARNING ! Always fit correct mattress type for intended mode of operation.
  • Page 15: Tilting The Bed

    Preparation Tilting the Bed Tilting the Bed Pull handle of locking mechanism toward front. Push handle down = head-up position. Pull handle up = head-down position. Release locking handle, the bed locks into position. Maximum tilt angles: Trendelenburg position: 20° Reverse Trendelenburg position: 15°...
  • Page 16: Tests Of Readiness For Operation

    Tests of Readiness for Operation Tests of Readiness for Operation Prior to initial operation Verify that supplied voltage agrees with the values indicated on the rating plate next to the power cable. Prior to each use Has the device been disinfected? Are the surrounding panels locked securely into place, and closed? Are all panels free of cracks and sharp edges?
  • Page 17: General Precautions During Patient Care

    The heating system provided with Babytherm 8000 WB is a mattress heater. Please refer to pages 26 - 28 for information about use of Babytherm 8000 WB in the presence of radiant heaters. Heater Safety Precautions WARNING ! Patient Temperature Monitoring Check the baby´s core temperature at regular...
  • Page 18: Checking Heater Readiness For Operation

    Heat Therapy Checking Heater Readiness for Operation WARNING ! Mattress Heater When mattress heating is used, the gel mattress must be placed directly on the heated tray surface. The Dräger gel mattress (part no. 2M 20 827) must not be used in any bassinets other than the Babytherm 8000.
  • Page 19 Checking Heater Readiness Heat Therapy for Operation 40 °C Babytherm 8000 ±1 °C Inop °C Check °C >37 °C <35 °C Mattress Heater Control Panel Power switch Button for extending range of settings to tempera- tures from 30 °C to 38.5 °C. Display of actual (measured) mattress temperature.
  • Page 20 Checking Heater Readiness Heat Therapy for Operation Switching heater unit on and activating self-test Babytherm 8000 Press power switch until it engages = ON. °C A microprocessor self-test now checks proper function of the unit. °C >36 °C >35 °C Check All LEDs light up;...
  • Page 21 Checking Heater Readiness Heat Therapy for Operation Checking LEDs, displays and audible alarm: Babytherm 8000 Press button: °C All LEDs light up for approximately 2 seconds; the numeric display shows 88.8, and an audible alarm is °C >36 °C generated. >35 °C Check After that, displays and LEDs go dark and the audible...
  • Page 22: Heater Operation

    Operation Operation without Concurrent Use of a Radiant Warmer Heater Operation Operation without Concurrent Use of a Radiant Warmer 37.0 °C 37,0°C Heater Operation Heat up mattress to 37 °C. Allow approximately 1 hour for thorough warm-up. During warm-up and in standby mode, cover the gel mattress with a blanket.
  • Page 23 5 °C three lines in the top of the display = the temperature is above 45 °C NOTE: This may occur e. g. when the Babytherm 8000 WB °C is stored in ambient conditions beyond the range of 5 to...
  • Page 24: Alarms

    Heat Therapy Heater Alarms Shut-Down Heater Alarms Deviation from set temperature Babytherm 8000 In case of deviations greater than ±1 °C between actual °C and set mattress temperature: The red LED ±1 °C flashes and an intermittent audible °C >36 °C >35 °C Check alarm is generated, and...
  • Page 25: Notes On Mattress Heater Settings

    Notes on Mattress Heater Settings Notes on settings assume use of the mattress heater without any additional radiant heat source. For information on use of the Babytherm 8000 WB in the presence of radiant heater systems, see pages 26 ff. WARNING ! Patient Temperature Monitoring Check the baby´s core temperature at regular...
  • Page 26: Operation With Concurrent Use Of A Radiant Heater

    The baby's core use with the Babytherm 8000 WB. Specifically, temperature must be monitored particularly combinations with radiant warmers that are not carefully during such treatments. The supply of...
  • Page 27 Operation with Concurrent Heat Therapy Use of a Radiant Heater Do not use the cover. Otherwise, a radiant heater will have no effect. Do not cover or dress patient. Using a radiant heater without Babytherm mattress heating Always use a foam rubber mattress! Do not use a gel mattress without mattress heating, or when mattress heating is switched off (see page 18)! 36.3 °C...
  • Page 28 Heat Therapy Operation with Concurrent Use of a Radiant Heater Using radiant heaters and mattress heating with a gel mattress To add warmth from below, in addition to a radiant 36.3 °C heater. 36,3°C Set mattress heater temperature to, e. g., 38 °C, set skin temperature at the radiant heater to, e.
  • Page 29: Taking X-Rays

    Taking X-Rays Place x-ray film cassette directly under patient (above mattress). Position the X-ray equipment above Babytherm 8000 WB. If a radiant heater is in use: Swivel radiant heater to the side or move radiant heater stand. After finishing with x-rays, swivel radiant heater back...
  • Page 30: Care

    Cleaning / Disinfecting / Sterilizing Care Cleaning / Disinfecting / Sterilizing The Babytherm 8000 WB heated bassinet must be CAUTION ! thoroughly cleaned and disinfected after each patient, or Certain components of the bassinet consist of ma- at least once a week. Perform any disinfection...
  • Page 31: Before Next Use

    CAUTION ! Do not launder mattresses in an automatic washing machine. Do not autoclave mattresses CAUTION ! Do not expose Babytherm 8000 WB to high level of ultraviolet radiation. It may cause cracks in the PMMA (perspex) parts. Before Next Use Reassemble components.
  • Page 32: Inspection And Maintenance

    Inspection and Replacement of Parts Maintenance Preventive Maintenance Intervals Inspection and Maintenance WARNING ! To avoid any risk of infection, clean and disinfect bassinet and accessories before any maintenance according to established hospital procedures – this applies also when returning units or parts for repair.
  • Page 33: Troubleshooting

    Troubleshooting Mattress Heater Problem/Message Likely Causes Remedy Red LED ±1 °C flashes, Measured mattress temperature During operation with radiant heater: actual value display flashes, deviates from set temperature reduce radiant heater output. by more than ±1 °C. intermittent audible alarm. Red LED 40 °C flashes, Mattress temperature >40 °C During operation with radiant heater:...
  • Page 34: What's What

    What is What Overview WARNING ! DANGER OF PATIENT INJURY NEVER LEAVE BABY UNATTENDED WHEN SURROUNDING PANELS ARE DOWN Overview Cover Chassis with 4 castors, 2 of which can be locked End panel Main WARNING/CAUTION label Side panel Accessory mounting rail Handle with inner unlocking handle for tilting the cot Inner rail with hook for drainage canister Mattress heater control panel...
  • Page 35: Mattress Heater Control Panel

    Mattress Heater What is What Control Panel 40 °C Babytherm 8000 ±1 °C Inop °C Check °C >37 °C <35 °C Mattress Heater Control Panel Power switch Button for extending range of settings to tempera- tures from 30 °C to 38.5 °C. Display of actual (measured) mattress temperature.
  • Page 36: Labels

    What is What Labels CAUTION ! KEEP TOP KEEP SPACE OF STORAGE ABOVE CABINET INCUBATOR CLEAR WHEN CLEAR WHEN LOWERING RAISING (both sides) Labels Main WARNING/CAUTION label DANGER ! RISK OF EXPLOSION IF USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS WARNING ! DISCONNECT SUPPLY BEFORE SERVICING REPAIRS ON THIS EQUIPMENT TO BE PERFORMED...
  • Page 37 Labels What is What Mattress heater WARNING label WARNING ! DO NOT SWITCH OFF MATTRESS HEATING WHILE A BABY IS LYING ON THE GEL MATTRESS AS THIS WILL RESULT IN COOLING OF THE PATIENT Mattress WARNING label WARNING ! USE RUBBER MATTRESS WHEN THE HEATING MATTRESS IS COLD.
  • Page 38: Technical Data

    Technical Data Ambient conditions For operation: 20 °C to 30 °C Temperature Atmospheric pressure 900 hPa to 1100 hPa Relative humidity 0 to 90%, no condensation In storage: –20 °C to 60 °C Temperature Atmospheric pressure 750 hPa to 1100 hPa Relative humidity 0 to 90%, no condensation Electrical safety...
  • Page 39 Technical Data Dimensions Length x Width 940 x 620 mm (37,0 x 24,4 inches) Total height (with cover) basic unit without height adjustment 1285 mm (50,6 inches) basic unit with height adjustment 1220 to 1515 mm (48,0 to 59,6 inches) Working height of the bed surface without height adjustment 950 mm (37,4 inches)
  • Page 40: Ordering Information

    Ordering Information Name and Description Part No. Name and Description Part No. Babytherm 8000 WB FR 00 050 Consumables Voltage level as ordered Foam rubber mattress 2M 21 012 To be specified with order: for unheated bassinet – Cover or no cover...
  • Page 41: Index

    Index larms..............24 echnical data............38 Ambient conditions............ 38 Temperature settings...........21, 28 Tests of readiness for operation......16, 18 efore next use............31 Tilting bed..............15 Bassinet, preparing........... 12 Troubleshooting............33 Bed, tilting..............15 ARNINGS and CAUTIONS, General....... 7 are................. 30 What's what............34 - 37 Checking, heater readiness for operation....18 Cover..............14, 26 -rays, taking............
  • Page 42: Notes

    Notes...
  • Page 43 Notes...
  • Page 44 This Operating Instruction Manual applies only to Babytherm 8000 WB (FR 00 050) or Babytherm 8000 OC without radiant warmer (FR 00 048) with Serial No.: Without entry of a Serial No. by Dräger, these Operating Instructions provide general information only and are not...

This manual is also suitable for:

Fr 00 050

Table of Contents