Dräger Babylog 8000 plus Operating Instructions Manual page 48

Infant care ventilator
Hide thumbs Also See for Babylog 8000 plus:
Table of Contents

Advertisement

Preparation
Checks of Readiness for Operation
Ventilator Checklist
Perform before each use of the ventilator.
After exchanging a patient circuit, only checkpoints
3 through 7 are required.
A copy of this checklist is provided as a writing board
and should be kept attached to the ventilator.
Check off each item on the check list using a
pencil. Date and initial.
Babylog 8000 plus
________________ ° Use chain to attach this checklist to ventilator
Serial No.
Operational verification procedure before each use.
What
How
1 Gas Supply
Connect high pressure Air and Oxygen hoses
at the back, insert probes into wall terminals
2 Patient System
Expiratory valve
Patient circuit
Water traps
Flow sensor connector
Connect test lung with tracheal tube CH 12
(2.5 mm ID) and ET-tube connector to wye.
3 Leak Test
Switch Babylog 8000 plus ON.
Press »Confirm« key,
Set ventilation mode CPAP:
Press keys »Vent. Mode«, »CPAP« and »On«.
Dial knobs »P
Press »Confirm« key,
Keep »man. Insp.« pressed:
48
Babylog 8000 plus Checklist
°
Always read Operating Manual first and observe
all instructions.
« to 80, »Insp. Flow * * * * « to 2,
insp
Date____________
________________
Signature
Requirement
Hose fittings: screws tightened
Connector probe: firmly inserted
firmly attached
complete
vertical position at lowest point
connected
Display:
Calibrate flow sensor!
Bar graph pressure display:
(80 ±2) cmH
O
2