Connecting Supplies
Gas Supply
WARNING !
Always use medical grade oxygen and air that
is dry and free from dust and oil. Contami-
nated gas may cause ventilator malfunction.
Supply pressures must be between 43.5 and 87 psi
(3 to 6 bar)
Screw high pressure air and oxygen hoses to sockets
on the back panel of Babylog and insert probes into
wall terminals.
CAUTION !
If Air and O
supply lines are contaminated with
2
particles or moisture, as is likely during the use
with a medical air compressor, use high pressure
line filters that retain at least 0.1 micron particles
and stop water droplets (e.g. Dräger part no.
D 800 130, D 800 132 ).
WARNING !
Always provide adequate ventilation!
When the equipment is being operated with
O
, ensure that the ambient concentration
2
does not exceed 24% O
of fire may be increased.
Electrical Supply
Line voltage must correspond to voltage range of
ventilator (see rating plate on back):
either : 220 V to 240 V
or
: 100 V to 127 V
Insert Babylog power plug into wall outlet.
WARNING !
To maintain grounding integrity, connect
only to a "hospital grade" receptacle.
Always disconnect supply before servicing.
. Otherwise, the risk
2