Operation
Special Functions
In order to maintain constant FiO2 :
Increase O
-concentration for the duration of the
2
aerosol treatment with the dial knob »O
using a value from the nomogram.
Example:
Insp. Flow *
= 10 L/min
O
-conc.-%
= 80 Vol.%
2
During aerosol application, set dial knob
»O
-Conc. %« to approximately 90 Vol.-%.
2
Ending nebulizer treatment
1 Retract ring on output port, connector will disengage.
Remove residual medication from nebulizer, unmount
nebulizer and process according to its Instructions for
Use.
Re-insert flow sensor element into flow sensor hous-
ing,
re-attach connector plug to flow sensor.
Install a cleaned and sterilized expiratory valve, see
page 29.
74
resulting FiO
resultierende FiO
[Vol.%]
100
-Conc. %«
2
90
80
70
60
50
40
30
20
1 2
2
2
100
90
80
70
60
50
40
30
21
4
6
8
10 12 14 16 18 20 22 24 26 28
Insp. Flow * [L/min]