5)
Confirm the zero calibration reading and then touch [Next].
6)
Touch [Target Conc.] and enter the target concentration of your span gas (ppm) followed by [Finish].
7)
When ready touch [Start] and apply your span gas to the sensor.
8)
Wait until the sensor reading
9)
Confirm that the gas reading is correct and touch [Finish]:
10) Remove the span gas from the sensor, and choose whether or not to calibrate another channel.
MAN0984_Iss 3_02/16
Pt. No. 3011M5001_3
Commissioning
becomes stable
.
72
Touchpoint Plus
Technical Handbook