5-18
IEEE-488 Operation
Table 5-7
Trigger command summary
Command
:ARM[:SEQuence[1]]
:LAYer2
:TCONfigure
:DIRection <name>
:DIRection?
:ASYNchronous
:ILINe <NRf>
:ILINe?
:OLINe <NRf>
:OLINe?
:TRIGger[:SEQuence[1]]
:IMMediate
:COUNt <n>
:AUTO <b>
:AUTO?
:COUNt?
:DELay <n>
:DELay?
:SOURce <name>
:SOURce?
:TIMer <n>
:TIMer?
:SIGNal
:TCONfigure
:PROTocol <name>
:PROTocol?
:DIRection <name>
:DIRection?
:ASYNchronous
:ILINe <NRf>
:ILINe?
:OLINe <NRf>
:OLINe?
:SSYNchronous
:LINE <NRf>
:LINE?
Notes:
1. Brackets ([ ]) are used to denote optional character sets. These optional characters do not have to be included in the
program message. Do not use brackets ([ ]) in the program message.
2. Angle brackets (< >) are used to indicate parameter type. Do not use angle brackets (< >) in the program message.
3. Upper case characters indicate the short-form version for each command word.
4. At least one space is required between a command word and the parameter.
Path to configure Triggers:
Enable (SOURce) or disable (ACCeptor) Bypass.
Query direction.
Path to configure asynchronous Trigger Link:
Select input line (1 to 6).
Query input line.
Select output line (1 to 6).
Query output line.
Path to program trigger layer:
Immediately perform a scan step.
Program measure count (1 to 9999, or INF).
Enable (1 or ON) or disable (0 or OFF) auto-count (scan-list-
length).
Query auto-count.
Query measure count.
Program delay (0 to 99999.999 seconds).
Query delay.
Select control source: HOLD, IMMediate, TIMer, MANual,
BUS, TLINk.
Query control source.
Set timer interval (0 to 99999.999 seconds).
Request the programmed timer interval.
Loop around control source.
Path to configure Triggers:
Select protocol: ASYNchronous, SSYNchronous.
Query protocol.
Enable (SOURce) or disable (ACCeptor) Bypass.
Query direction.
Path to configure asynchronous Trigger Link:
Select input line (1 to 6).
Query input line.
Select output line (1 to 6).
Query output line.
Path to configure semi-synchronous Trigger Link:
Select trigger line (1 to 6).
Query trigger line.
Return to