9.4

Mounting support

After performing any kind of assembling work, please make sure all the grounded contacts are
connected correctly!
9.4.1

Replacing the front housing, the incremental shaft encoder (potentiometer) and the display

Replacing the front housing
• Remove the lower covering by turn out the 2 screws (see figure 40).
• Pull the lower edge of the covering away from the mounting plate.
• Pull the covering out of the upper guiding rail.
• Loosen the 5 screws under the control panel and pull away the panel forward. Use therefore
the pictured tools.
Remove the interface cable on the control board. (connection between control board and control
panel).