Variable screens

The Rosemount
Table 6-1. LCD Display Variables
Parameter
Level
Distance
Level Rate
Signal Strength
Volume
Electronics Temperature
Signal Quality
Scaled Variable
Percent of Range
Primary Variable
Auxiliary Percent of
Range
User Defined Variable
1.
Only for transmitters ordered with Smart Diagnostics Suite (option code DA1).
2.
Default, user selectable display text (up to five characters).
6.2

Set up the LCD display

It is possible to specify the variables to be presented on the optional LCD display.
Rosemount Radar Master Plus
1. Under Configure, select Device Setup, and then select the Display tab.
2. Select the desired variables to be displayed on the LCD display.
3. Select Save.
AMS Device Manager and Field Communicator
1. Select Configure > Manual Setup > Device Setup > Display.
2. Select the desired variables to be displayed on the LCD display.
3. Select Send.
76
5408 and 5408:SIS Level Transmitters can display the following variables:
Presentation
on display
LEVEL
DIST
LR
AMP
VOLUM
ITEMP
SIG
(1)
QUALITY
(1)
(2)
SCALE
PV
%RANGE
AUX
%RANGE
(1)
(2)
USER
Description
The current level measurement value.
Distance from the upper reference point to the product surface.
The current velocity at which the level is moving. A positive value
indicates the surface is moving up.
The signal amplitude of the surface echo.
Volume of the product at the current level.
The current temperature at the electronics.
The quality of product surface echo signal compared to surface
threshold and noise.
A variable calculated from a scaling table (as defined by pairs of
input/scaled values).
A variable value expressed in percent within a range defined by a
Lower Range Value (LRV) and an Upper Range Value (URV).
A variable value expressed in percent within a range defined by a
Lower Range Value (LRV) and an Upper Range Value (URV).
A variable associated with a selected register in the device. Refer
Table C-4 on page 200
for a list of suitable register variables.