Procedure
1. Remove the lock nut.
2. Mount the O-ring.
3. Mount the antenna adapter on flange/manhole cover.
Ensure the antenna adapter fits tightly to the flange/manhole cover.
Mechanical Installation
E
F
G
> 0.59 in. (15 mm)
Ø 3.98 ± 0.02 in.
(Ø 101 ± 0.6 mm)
OR
BSPP (G) 3½-in.
Max. 1.18 in. (30 mm)